Description
HI5522 Multiparameter
HI5522 is a RESEARCH GRADE ISE/pH/ORP/°C – EC/TDS/Salinity -with 2 Channel Benchtop
The HI5522 is a multi parameter and benchtop meter can ,measure pH/mV/ISE/EC/TDS/Salinity/Resistivity
please see important items of HI5522 at a glance :
The HI5522 has Highly Customizable User Interface .
The HI5522 has Color Graphic LCD.
Capacitive Touch .
Four Ring Conductivity Probe .
Choice of Calibration …
CAL Check™ .
ISE Measurement with Choice of Concentration Units .
ISE Measurement with Incremental Methods .
Data Logging.
Data Transfer.
Contextual Help .
To know HI5522 HANNA price or buy HI5522 Multiparameter , please contact Setare Arsh Aria company.
pH Specifications
pH RangepH ResolutionpH Accuracy (@25ºC/77ºF)pH CalibrationpH Temperature CompensationmV RangemV ResolutionmV AccuracyRelative mV Offset Range
-2.000 to 20.000 pH |
0.1, 0.01, 0.001 pH |
±0.1 pH, ±0.01 pH, ±0.002 pH ±1 LSD |
automatic, up to five point calibration, eight standard buffers available (1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01,12.45) and five custom buffers |
automatic or manual from -20.0 to 120.0 °C |
±2000 mV |
0.1 mV |
±0.2 mV ±1 LSD |
±2000 mV |
EC Specifications
EC RangeEC ResolutionEC Accuracy (@25ºC/77ºF)EC Calibration
0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC* |
0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm |
±1% of reading (±0.01 μS/cm) |
automatic standard recognition (0.000 μS/cm, 84.00 μS/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration |
TDS Specifications
TDS RangeTDS ResolutionTDS Accuracy (@25ºC/77ºF)
0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor) |
0.001 ppm, 0.01 ppm, 0.1 ppm, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt |
±1% of reading (±0.01 ppm) |
Resistivity Specifications
Resistivity RangeResistivity ResolutionResistivity Accuracy
1.0 to 99.9 Ω•cm; 100 to 999 Ω•cm; 1.00 to 9.99 KΩ•cm; 10.0 to 99.9 KΩ•cm; 100 to 999 KΩ•cm; 1.00 to 9.99 MΩ•cm; 10.0 to 100.0 MΩ•cm |
0.1 Ω•cm; 1 Ω•cm; 0.01 KΩ•cm; 0.1 KΩ•cm; 1 KΩ•cm; 0.01 MΩ•cm; 0.1 MΩ•cm* |
±2% of reading (±1 Ω•cm) |
Salinity Specifications
Salinity RangeSalinity ResolutionSalinity Accuracy (@25ºC/77ºF)Salinity Calibration
practical scale: 0.00 to 42.00 psu; natural seawater scale: 0.00 to 80.00 ppt; percent scale: 0.0 to 400.0% |
0.01 for practical scale/natural seawater scale; 0.1% for percent scale |
±1% of reading |
percent scale—one-point (with HI7037 standard) |
Temperature Specifications
Temperature RangeTemperature ResolutionTemperature AccuracyTemperature CompensationTemperature CoefficientReference Temperature
-20.0 to 120.0 ºC**; -4.0 to 248.0 ºF**; 253.15 to 393.15 K** |
0.1 °C, 0.1 °F, 0.1 K |
±0.2 °C; ±0.4 °F; ±0.2 K (without probe) |
disabled, linear and non-linear (natural water) |
0.00 to 10.00 %/°C |
5.0 to 30.0 °C |
ISE Specifications
ISE RangeISE ResolutionISE Accuracy @ 25 °C/77 °FISE Calibration Points
1 x 10⁻⁶ to 9.99 x 10¹⁰ concentration |
1; 0.1; 0.01; 0.001 concentration |
±0.5% (monovalent ions); ±1% (divalent ions) |
automatic, up to five point calibration, five fixed standard solutions available (0.1, 1, 10, 100, 1000 choice of concentration) and five user defined units |
Additional Specifications
Cell Constant Cell Type Electrode/Probe Temperature Probe ProfilesUSP ⟨645⟩ Compliant GLPLoggingInput Channels DisplayConnectivityEnvironmentPower Supply DimensionsWeightOrdering Information Notes
0.0500 to 200.00 |
4 cells |
HI1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included); HI76312 platinum, four-ring EC/TDS probe with and 1 m (3.3’) cable (included) |
HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) |
up to 10, 5 each channel |
yes |
calibration data including date, time, buffers used, offset and slope for pH. Cell constant, reference temperature, temperature coefficient, calibration points, calibration time stamp, probe offset for conductivity |
record: 100,000 data point storage, up to 100 lots with max. 50,000 records/lot; interval: settable between 1 second and max log time of 180 minutes; type: automatic, manual, AutoHOLD; additional: 200 records USP |
1 pH/ORP/ISE + 1 EC/TDS/Salinity/Resistivity |
color graphic LCD with on-screen help, graphing, and custom color configuration |
USB |
0 to 50°C (32 to 122°F; 273 to 323 K), RH max 95% non-condensing |
12 VDC adapter (included) |
160 x 231 x 94 mm (6.3 x 9.1 x 3.7”) |
1.2 kg (2.64 lbs.) |
HI5522 is supplied with HI1131 pH electrode, HI7662-T temperature probe, HI76312 conductivity probe, HI76404W electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachet (2), HI7082 3.5M KCl electrolyte solution (30 mL), 12 VDC power adapter and instructions. |
*Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation. **Reduced to actual probe limits |