Description
HI83300 Multiparameter Photometer
HI83300 Multiparameter Photometer
HI83300 is a benchtop photometer and useful device for most of laboratories whom are looking to measure most of water parameters .
HI83300 Multiparameter Photometer
HI83300 is multiparameter photomer and Ph meter that can measure 37 key water quality parameters via optical method.
HI83300 is ex termly small . its similar to portable and bench top device both.
below is some characters of HI83300:
- High Efficiency LED Light Source
- High Quality Narrow Band Interference Filters
- Reference Detector for a Stable Light Source
- Large Cuvette Size
- Focusing Lens for Greater Light Yield
Key features OF hi83300 been listed :
- Advanced optical system with brighter, long-lasting LED light source
- Backlit 128 x 64 Pixel Graphic Liquid Crystal Display
- Built-in Reaction Timer for Photometric Measurements
- Absorbance mode – concentration versus absorbance
- Units of Measurement plus chemical form displayed
- Result conversion at the touch of a button
- Cuvette cover
- pH and temperature measurement with a single probe
- Good Laboratory Practice (GLP) – track calibration information including date, time, buffers used, offset and slope for traceability
- CAL Check alerts user to potential problems during the calibration process
- Data Logging – Up to 1000 photometric and pH readings can be stored
- Logged readings can be quickly and easily transferred to a flash drive or to a PC. Data is exported as a .CSV file for use with spreadsheet programs
- Battery Status Indicator
- Error Messages
HI83300 Multiparameter Photometer
HI83300 Multiparameter Photometer
HI83300 Multiparameter Photometer
To know HI83300 HANNA price or buy HI83300 Multiparameter photometer , please contact Setare Arsh Aria company.
Specifications
Specifications list
Specifications
GENERAL |
|
|
MEASUREMENT CHANNELS |
|
5 X OPTICAL CHANNELS; 1 X SMART ELECTRODE CHANNEL (pH MEASUREMENT) |
Absorbance Range |
|
0.000Abs-4.000Abs |
Absorbance Resolution |
|
0.001 Abs |
Absorbance Accuracy |
|
±0.003 Abs (at 1.000 Abs) |
Absorbance Light Source |
|
Light-emitting diode |
Absorbance Bandpass Filter Bandwidth |
|
8nm |
Absorbance Bandpass Filter Wavelength Accuracy |
|
±1.0nm |
Light Detector |
|
Silicon photocell |
Cuvette Type |
|
round, 24.6 mm diameter |
Number of Methods |
|
128 max |
pH Range |
|
-2.00 to 16.00 pH (±1000mV)* |
pH Resolution |
|
0.01 pH (0.01 mV)* |
Temperature Compensation |
|
ATC (-5.0 to 100.0°C;*) |
Additional Specifications: |
|
|
Logging |
|
1000 readings (mixed photometer and electrode) |
Display |
|
128 x 64 pixel B/W LCD with backlight |
USB-A (host) functions |
|
mass-storage host |
USB-B ( Device) Functions |
|
power input, mass-storage device |
Battery Life |
|
>500 Photometer measurements, or 50 hours of continuous pH measurement |
Power supply |
|
5 VDC USB 2.0 power adapter/ type micro-B connector 3.7 VDC Li-polymer rechargeable battery, non-serviceable |
Environment |
|
0 to 50°C; 0 to 95% RH, non-condensing |
Dimensions |
|
206 x 177 x 97mm |
Weight |
|
1.0kg |
Ordering Information |
|
HI-83300 is supplied with sample cuvettes and caps (x4), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual. |
*Limits will be reduced to actual sensor limits |
|
|
ABSORBANCE |
|
|
Absorbance Range |
|
0.000 to 4.000 Abs |
Absorbance Resolution |
|
0.001 Abs |
Absorbance Accuracy |
|
+/-0.003Abs @ 1.000 Abs |
ALKALINITY |
|
Reagents – HI-93755-01 (100 tests); HI-93755-03 (300 tests) |
Alkalinity Range |
|
Freshwater: 0 to 500 mg/L (as CaCO3) Marine: 0 to 300 mg/L (as CaCO3) |
Alkalinity Resolution |
|
1 mg/L |
Alkalinity Accuracy |
|
±5 mg/L ±5% of reading |
Alkalinity Method |
|
Colorimetric method |
ALUMINIUM |
|
Reagents – HI-93712-01 (100 tests); HI-93712-03 (300 tests) |
Aluminium Range |
|
0.00 to 1.00 mg/L (as Al3+) |
Aluminium Resolution |
|
0.01 mg/L |
Aluminium Accuracy |
|
±0.04 mg/L ±4% of reading |
Aluminium Method |
|
Adaptation of the aluminon method |
AMMONIA |
|
Reagents: HR – HI-93733-01 (100 tests); HI-93733-03 (300 tests)
MR – HI-93715-01 (100 tests); HI-93715-03 (300 tests)
LR – HI-93700-01 (100 tests); HI-93700-03 (300 tests) |
Ammonia Range |
|
Low Range: 0.00 to 3.00 mg/L (as NH3-N) Medium Range: 0.00 to 10.00 mg/L (as NH3-N) High Range: 0.0 to 100.0 mg/L (as NH3-N) |
Ammonia Resolution |
|
0.01 mg/L; 0.1 mg/L |
Ammonia Accuracy |
|
Low Range: ±0.04 mg/L ±4% of reading Medium Range: ±0.05 mg/L ±5% of reading High range: ±0.5 mg/L ±5% of reading at 25 °C |
Ammonia Method |
|
Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method |
ANIONIC SURFACTANTS |
|
|
Anionic Surfactants Range |
|
0.00 to 3.50 mg/L (as SDBS) |
Anionic Surfactants Resolution |
|
0.01 mg/L |
Anionic Surfactants Accuracy |
|
±0.04 mg/L ±3% of reading |
Anionic Surfactants Method |
|
Adaptation of the USEPA method 425.1 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 5540C, Anionic Surfactants as MBAS. |
BROMINE |
|
Reagents: HI-93716-01 (100 tests); HI-93716-03 (300 tests) |
Bromine Range |
|
0.00 to 8.00 mg/L (as Br2) |
Bromine Resolution |
|
0.01 mg/L |
Bromine Accuracy |
|
±0.08 mg/L ±3% of reading |
Bromine Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. |
CALCIUM |
|
Reagents: HI-937521-01 (50 tests); HI-937521-03 (150 tests) |
Calcium Range |
|
Freshwater: 0 to 400 mg/L (as Ca2+) Seawater: 200 to 600 mg/L (as Ca2+) |
Calcium Resolution |
|
1 mg/L |
Calcium Accuracy |
|
Freshwater: ±10 mg/L ±5% of reading Seawater: ±6% of reading |
Calcium Method |
|
Freshwater: Adaptation of the Oxalate method Seawater: Adaptation of the Zincon method |
CHLORIDE |
|
Reagents: HI-93753-01 (100 tests); HI-93753-03 (300 Tests) |
Chloride Range |
|
0.0 to 20.0 mg/L (as Cl⁻) |
Chloride Resolution |
|
0.1 mg/L |
Chloride Accuracy |
|
±0.5 mg/L ±6% of reading at 25 °C |
CHLORINE DIOXIDE |
|
Reagents: HI-93738-01 (100 tests); HI-93738-03 (300 tests) |
Chlorine Dioxide Range |
|
0.00 to 2.00 mg/L (as ClO2) |
Chlorine Dioxide Resolution |
|
0.01 mg/L |
Chlorine Dioxide Accuracy |
|
±0.10 mg/L ±5% of reading |
Chlorine Dioxide Method |
|
Adaptation of the Chlorophenol Red method. |
CHLORINE |
|
Free Chlorine Reagents: HI-93701-01 (100 tests); HI-93701-03 (300 tests); HI-93701-F (liquid) (300 tests)
Chlorine UHR Reagents: HI-95771-01 (100 tests); HI-95771-03 (300 tests)
Chlorine, Free and Total HR: HI-93734-01 (100 tests); HI-93734-03 (300 tests)
Chlorine, Free ULR: HI-95762-01 (100 tests); HI-95762-03 (300 tests)
Chlorine Total: HI-93711-01 (100 tests); HI-93711-03 (300 tests); HI-93711-T (300 tests liquid)
Chlorine, Total ULR: HI-95761-01 (100 tests); HI-95761-03 (300 tests) |
Free Chlorine Range |
|
0.00 to 5.00 mg/L (as Cl2) Ultra Low Range: 0.000 to 0.500 mg/L (as Cl2) |
Free Chlorine Resolution |
|
0.01 mg/L Ultra Low Range: 0.001 mg/L |
Free Chlorine Accuracy |
|
±0.03 mg/L ±3% of reading Ultra Low Range: ±0.020 mg/L ±3% of reading |
Total Chlorine Range |
|
0.00 to 5.00 mg/L (as Cl2) Ultra Low Range: 0.000 to 0.500 mg/L (as Cl2) Ultra High Range: 0 to 500 mg/L (as Cl2) |
Total Chlorine Resolution |
|
0.01 mg/L; 0.001 mg/L; 1 mg/L |
Total Chlorine Accuracy |
|
±0.03 mg/L ±3% of reading Ultra Low Range: ±0.020 mg/L ±3% of reading Ultra High Range: ±3 mg/L ±3% of reading |
Chlorine Method |
|
Adaptation of the EPA 330.5 DPD method Free Chlorine (ULR) & Total Chlorine (UHR): Adaptation of the Standard Methods for Examination of Water and Wastewater, 20th edition, 4500-Cl |
CHROMIUM, HEXAVALENT |
|
Reagents: HR – HI-93723-01 (100 tests); HI-93723-03 (300 tests)
LR – HI-93749-01 (100 tests); HI-93749-03 (300 tests) |
Chromium, Hexavalent Range |
|
Low Range: 0 to 300 μg/L (as Cr6+) High Range: 0 to 1000 μg/L (as Cr6+) |
Chromium, Hexavalent Resolution |
|
1 μg/L |
Chromium, Hexavalent Accuracy |
|
Low Range: ±1 μg/L ±4% of reading High Range: ±5 μg/L ±4% of reading |
Chromium, Hexavalent Method |
|
Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. |
COLOUR, WATER |
|
|
Colour, Water Range |
|
0 to 500 PCU (Platinum Cobalt Units) |
Colour, Water Resolution |
|
1 PCU |
Colour, Water Accuracy |
|
±10 PCU ±5% of reading |
Colour, Water Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method. |
COPPER |
|
Reagents: HR – HI-93702-01 (100 tests); HI-93702-03 (300 tests)
Total copper – HI-93702T – 01 (100 tests); HI-93702T-03 (300 tests)
LR – HI-95747-01 (100 tests); HI-95747-03 (300 tests) |
Copper Range |
|
Low Range: 0.000 to 1.500 mg/L (as Cu2+) High range: 0.00 to 5.00 mg/L (as Cu2+) |
Copper Resolution |
|
0.001 mg/L; 0.01 mg/L |
Copper Accuracy |
|
Low Range: ±0.01 mg/L ±5% of reading High Range ±0.02 mg/L ±4% of reading |
Copper Method |
|
Adaptation of the EPA bicinchoninate method |
CYANURIC ACID |
|
Reagents: HI-93722-01 (100 tests); HI-93722-03 (300 tests) |
Cyanuric Acid Range |
|
0 to 80 mg/L (as CYA) |
Cyanuric Acid Resolution |
|
1 mg/L |
Cyanuric Acid Accuracy |
|
±1 mg/L ±15% of reading |
Cyanuric Acid Method |
|
Adaptation of the turbidimetric method |
FLUORIDE |
|
Reagents: HR – HI-93739-01 (100 tests); HI-93739-03 (300 tests)
LR – HI-93729-01 (100 tests); HI-93729-03 (300 tests) |
Fluoride Range |
|
Low Range: 0.00 to 2.00 mg/L (as F-) High range: 0.0 to 20.0 mg/L (as F-) |
Fluoride Resolution |
|
0.01 mg/L; 0.1 mg/L |
Fluoride Accuracy |
|
Low Range: ±0.03 mg/L ±3% of reading High Range: ±0.5 mg/L ±3% of reading |
Fluoride Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, SPADNS method |
HARDNESS, CALCIUM |
|
Reagents: HI-93720-01 (100 tests); HI-93720-03 (300 tests) |
Hardness, Calcium Range |
|
0.00 to 2.70 mg/L (as CaCO3) |
Hardness, Calcium Resolution |
|
0.01 mg/L |
Hardness, Calcium Accuracy |
|
±0.11 mg/L ±5% of reading |
Hardness, Calcium Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method |
HARDNESS, MAGNESIUM |
|
Reagents: HI-93719-01 (100 tests); HI-93719-03 (300 tests) |
Hardness, Magnesium Range |
|
0.00 to 2.00 mg/L (as CaCO3) |
Hardness, Magnesium Resolution |
|
0.01 mg/L |
Hardness, Magnesium Accuracy |
|
±0.11 mg/L ±5% of reading |
Hardness, Magnesium Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA Colorimetric method |
HARDNESS, TOTAL |
|
Reagents: HR – HI-93735-02 (100 tests); MR – HI-93735-01 (100 tests); LR – HI-93735-00 (100 tests); HR MR &LR – HI-93735-0 (100 x 3 tests) |
Hardness, Total Range |
|
Low Range: 0 to 250 mg/L (as CaCO3) Medium Range: 200 to 500 mg/L (as CaCO3) High Range: 400 to 750 mg/L (as CaCO3) |
Hardness, Total Resolution |
|
1 mg/L |
Hardness, Total Accuracy |
|
Low Range: ±5 mg/L ±4% of reading Medium Range: ±7 mg/L ±3% of reading High Range: ±10 mg/L ±2% of reading |
Hardness, Total Method |
|
Adaptation of the EPA recommended method 130.1 |
HYDRAZINE |
|
Reagents: HI-93704-01 (100 tests); HI-93704-03 (300 tests) |
Hydrazine Range |
|
0 to 400 μg/L (as N2H4) |
Hydrazine Resolution |
|
1 μg/L |
Hydrazine Accuracy |
|
±4% of full scale reading |
Hydrazine Method |
|
Adaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method |
IODINE |
|
Reagents: HI-93718-01 (100 tests); HI-93718-03 (300 tests) |
Iodine Range |
|
0.0 to 12.5 mg/L (as I2) |
Iodine Resolution |
|
0.1 mg/L |
Iodine Accuracy |
|
±0.1 mg/L ±5% of reading |
Iodine Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method |
IRON |
|
Reagents: HR – HI-93721-01 (100 tests); HI-93721-03 (300 tests)
LR – HI-93746-01 (50 tests); HI-93746-03 (150 tests) |
Iron Range |
|
Low Range: 0.000 to 1.600 mg/L (as Fe) High Range: 0.00 to 5.00 mg/L (as Fe) |
Iron Resolution |
|
0.001 mg/L; 0.01 mg/L |
Iron Accuracy |
|
Low Range: ±0.01 mg/L ±8% of reading High Range: ±0.04 mg/L ±2% of reading |
Iron Method |
|
Low Range: Adaptation of the TPTZ Method High Range: Adaptation of the EPA Phenanthroline method 315B, for natural and treated waters |
MAGNESIUM |
|
Reagents: HI-937520-01 (50 tests); HI-937520-03 (150 tests) |
Magnesium Range |
|
0 to 150 mg/L (as Mg2+) |
Magnesium Resolution |
|
1 mg/L |
Magnesium Accuracy |
|
±5 mg/L ±3% of reading |
Magnesium Method |
|
Adaptation of the Calmagite method |
MANGANESE |
|
Reagents: HR – HI-93709-01 (100 tests); HI-93709-03 (300 tests)
LR – HI-93748-01 (50 tests); HI-93748-03 (150 tests) |
Manganese Range |
|
Low Range: 0 to 300 µg/L (as Mn) High range: 0.0 to 20.0 mg/L (as Mn) |
Manganese Resolution |
|
1 µg/L; 0.1 mg/L |
Manganese Accuracy |
|
Low Range: ±0.02 mg/L ±3% of reading High Range: ±0.3 mg/L ±5% of reading |
Manganese Method |
|
Low Range: Adaptation of the PAN Method High Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Periodate method |
MOLYBDENUM |
|
Reagents: HI-93730-01 (100 tests); HI-93730-03 (300 tests) |
Molybdenum Range |
|
0.0 to 40.0 mg/L (as Mo6+) |
Molybdenum Resolution |
|
0.1 mg/L |
Molybdenum Accuracy |
|
±0.3 mg/L ±5% of reading |
Molybdenum Method |
|
Adaptation of the mercaptoacetic acid method |
NICKEL |
|
Reagents: HR – HI-93726-01 (100 tests); HI-93726-03 (300 tests)
LR – HI-93740-01 (50 tests); HI-93740-03 (150 tests) |
Nickel Range |
|
Low Range: 0.000 to 1.000 mg/L (as Ni) High Range: 0.00 to 7.00 g/L (as Ni) |
Nickel Resolution |
|
0.001 mg/L; 0.01 g/L |
Nickel Accuracy |
|
Low range: ±0.010 mg/L ±7% of reading High Range: ±0.07 g/L ±4% of reading |
Nickel Method |
|
Low Range: Adaptation of the PAN method High Range: Adaptation of the photometric method |
NITRATE |
|
Reagents: HI-93728-01 (100 tests); HI-93728-03 (300 tests) |
Nitrate Range |
|
0.0 to 30.0 mg/L (as NO3– N) |
Nitrate Resolution |
|
0.1 mg/L |
Nitrate Accuracy |
|
±0.5 mg/L ±10% of reading |
Nitrate Method |
|
Adaptation of the cadmium reduction method |
NITRITE |
|
Reagents: HR – HI-93708-01 (100 tests); HI-93708-03 (300 tests)
LR – HI-93707-01 (100 tests); HI-93707-03 (300 tests) |
Nitrite Range |
|
Freshwater Low Range: 0 to 600 μg/L (as NO2–N) High Range: 0 to 150 mg/L (as NO2-) Seawater Ultra Low Range: 0 to 200 μg/L (as NO2–N) |
Nitrite Resolution |
|
Freshwater: 1 μg/L; 1 mg/L Seawater: 1 μg/L |
Nitrite Accuracy |
|
Freshwater Low Range: ±20 μg/L ±4% of reading High Range: ±4 mg/L ±4% of reading Seawater ±10 μg/L ±4% of reading |
Nitrite Method |
|
Adaptation of the EPA Diazotization method 354.1 |
OXYGEN, DISSOLVED |
|
Reagents: HI-93732-01 (100 tests); HI-93732-03 (300 tests) |
Oxygen, Dissolved Range |
|
0.0 to 10.0 mg/L (as O2) |
Oxygen, Dissolved Resolution |
|
0.1 mg/L |
Oxygen, Dissolved Accuracy |
|
±0.4 mg/L ±3% of reading |
Oxygen, Dissolved Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method |
OXYGEN, SCAVENGER |
|
|
Oxygen, Scavenger Range |
|
0 to 1000 μg/L (as DEHA) 0.00 to 1.50 mg/L (as Carbohydrazide) 0.00 to 2.50 mg/L (as Hydroquinone) 0.00 to 4.50 mg/L (as ISO-ascorbic acid) |
Oxygen, Scavenger Resolution |
|
1 μg/L (DEHA); 0.01 mg/L |
Oxygen, Scavenger Accuracy |
|
±5 μg/L ±5% of reading (DEHA) |
Oxygen, Scavenger Method |
|
Adaptation of the iron reduction method |
OZONE |
|
Reagents: HI-93757-01 (100 tests); HI-93757-03 (300 tests) |
Ozone Range |
|
0.00 to 2.00 mg/L (as O3) |
Ozone Resolution |
|
0.01 mg/L |
Ozone Accuracy |
|
±0.02 mg/L ±3% of reading |
Ozone Method |
|
Colorimetric DPD Method |
pH |
|
Reagents: HI-93710-01 (100 tests); HI-93710-03 (300 tests) |
pH Range |
|
Photometer: 6.5 to 8.5 pH pH electrode: -2.00 to 16.00 pH |
pH Resolution |
|
Photometer: 0.1 pH pH electrode: 0.1 pH |
pH Accuracy |
|
Photometer: ±0.1 pH pH electrode: ±0.01 pH |
pH Calibration |
|
Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
pH Temperature Compensation |
|
Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used |
pH CAL Check (electrode diagnostics) |
|
clean electrode and check buffer/check probe displayed during calibration |
pH Method |
|
Photometer: phenol red |
pH-mV Range |
|
±1000 mV |
pH-mV Resolution |
|
0.1 mV |
pH-mV Accuracy |
|
±0.2 mV |
PHOSPHATE |
|
Reagents: HR – HI-93717-01 (100 tests); HI-93717-03 (300 tests)
LR – HI-93713-01 (100 tests); HI-93713-03 (300 tests) |
Phosphate Range |
|
Freshwater Low Range: 0.00 to 2.50 mg/L (as PO43-) High range: 0.0 to 30.0 mg/L (as PO43-) Seawater Ultra Low Range: 0 to 200 μg/L (as P) |
Phosphate Resolution |
|
Freshwater: 0.01 mg/L; 0.1 mg/L Seawater: 1 μg/L |
Phosphate Accuracy |
|
Freshwater Low Range: ±0.04 mg/L ±4% of reading High Range: ±1 mg/L ±4% of reading Seawater Ultra Low Range: ±5 μg/L ±5% of reading |
Phosphate Method |
|
Freshwater Low Range: Adaptation of the Ascorbic Acid method Freshwater High Range and Seawater Ultra Low Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method |
POTASSIUM |
|
Reagents: HI-93750-01 (100 tests); HI-93750-03 (300 tests) |
Potassium Range |
|
0.0 to 20.0 mg/L (as K) |
Potassium Resolution |
|
0.1 mg/L |
Potassium Accuracy |
|
±3.0 mg/L ±7% of reading |
Potassium Method |
|
Adaptation of the Turbidimetric Tetraphenylborate method |
SILICA |
|
Reagents: HR – HI-96770-01 (100 tests); HI-96770-03 (300 tests)
LR – HI-93705-01 (100 tests); HI-93705-03 (300 tests) |
Silica Range |
|
Low Range: 0.00 to 2.00 mg/L (as SiO2) High Range: 0 to 200 mg/L (as SiO2) |
Silica Resolution |
|
0.01 mg/L; 1 mg/L |
Silica Accuracy |
|
Low Range: ±0.03 mg/L ±3% of reading High Range: ±1 mg/L ±5% of reading |
Silica Method |
|
Low Range: Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue method High Range: Adaptation of the USEPA Method 370.1 and Standard Method 4500-SiO2 |
SILVER |
|
Reagents: HI-93737-01 (50 tests); HI-93737-03 (150 tests) |
Silver Range |
|
0.000 to 1.000 mg/L (as Ag) |
Silver Resolution |
|
0.001 mg/L |
Silver Accuracy |
|
±0.020 mg/L ±5% of reading |
Silver Method |
|
Adaptation of the PAN method |
SULPHATE |
|
Reagents: HI-93751-01 (100 tests); HI-93751-03 (300 tests) |
Sulphate Range |
|
0 to 150 mg/L (as SO42-) |
Sulphate Resolution |
|
1 mg/L |
Sulphate Accuracy |
|
±5 mg/L ±3% of reading |
Sulphate Method |
|
Turbidimetric – Sulphate is precipitated with barium chloride crystals |
ZINC |
|
Reagents: HI-93731-01 (100 tests); HI-93731-03 (300 tests) |
Zinc Range |
|
0.00 to 3.00 mg/L (as Zn) |
Zinc Resolution |
|
0.01 mg/L |
Zinc Accuracy |
|
±0.03 mg/L ±3% of reading |
Zinc Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method |