Description
HI934 coulometric Karl Fischer
HI934 is new design of coulometric Karl Fischers .
This type of Karl Fisher measureS the water amount of products from 1ppm to 5%.
HI934 has very small size that is suitable for small laboratories.
The main details of HI934 been listed as below:
- Narrow footprint, compact design, little space required
- Housing made of robust, chemical-resistant plastic
- Integrated and powerful algorithms for the determination of endpoint and absolute / relative drift
- Averaging for titrant standardization and sample analysis
- Minimized water vapor ingress through sealed measuring room
- Car interface for automatic weighing
- Storage of up to 100 titration methods
- customizable reports
- Display of warning and error messages in plain text
To know HI934 HANNA price or buy HI934 coulometric Karl Fischer , please contact Setare Arsh Aria company.
Technical specifications:
Measuring range: | 1 ppm to 5% |
Resolution: | 0.1 ppm |
Result unit: | %, ppm, mg / g, µg / g, mg, µg, mg / mL, µg / mL, ppt, mgBr / 100g, gBr / 100g, mgBr, gBr |
Type of sample: | liquid and solid (external dissolution or extraction) |
Pre-titration: | automatically |
Drift correction: | automatic or custom value |
Endpoint Criteria: | fixed mV persistence, relative or absolute drift stop |
Dosage: | dynamic with 3 speed levels |
Result statistics: | Mean and standard deviation |
Titration vessel: | Made of borosilicate glass with standard cone glass joint connections, volume from 100 to 200 mL, septum made of silicone rubber, septum cap thread GL-18 with standard Kegel19 reagent port |
Electrode type: | Detector & generator electrode |
Magnetic stirrer: | magnetic, optically regulated, digital | Speed: 200 to 2000 rpm | Resolution: 100 rpm |
external stirrer: | 4-pin mini-DIN connector enables control of an external agitator |
Storage method: | up to 100 (standard and custom) methods |
Storage reports: | up to 100 full titration reports and drift rate reports |
Reagent system: | Sealed system with integrated membrane air pump |
Display: | 5.7 ” graphic color display with backlight |
Peripherals: | PC (USB standard B); Storage medium (USB standard A); Analytical balance (DB-9 connection); Printer (DB-25 connector); Keyboard (6-pin mini DIN) |
Languages: | English, Portuguese, Spanish and French |
Power supply / power connection: | 100-240 Vac., 50/60 Hz / 0.5 Amps |
Housing material: | ABS / PC and steel |
Keyboard: | polyester |
Operating conditions: | 10 to 40 ° C; Max. 80% rel. humidity |
Environmental conditions: | -20 to 70 ° C, rel. Humidity max. 95% |
Dimensions: | 315 x 205 x 375 mm |
Weight: | Max. 4.3kg (with a pump, stirrer and sensors) |
Reviews
There are no reviews yet.