Description
HI5321 EC / TDS / Salinity and Temperature
Product information “HI5321 RESEARCH GRADE EC / TDS / Salinity and Temperature with Color-Display”
HI5321 is EC / TDS / Salinity and Temperature bech top meter made from HANNA.
HI5321 is professional bench top meter that is top choice for research institute.
HI5321 is supplied with HI76312 four ring probe.
HI5321 is supplied with the HI76312 four-ring conductivity probe.
HI76312 probe measure conductivity in wide range from 0.000 μS/cm to 1000.0 mS/cm*.
HI5321 has GLP capability.
HI5321 features a color graphic LCD with on-screen help, graphic, and custom color configurations.
HI5321 is very top choice for pharmaceutical industry.
HI5321 works with selectable loggin modes.automatic, manual and AutoHold logging
some HI5321 featurs been listed as below :
- Highly Customizable User Interface –
- Color Graphic LCD
- Capacitive Touch Four Ring Conductivity Probe Choice of Calibration
- GLP Data
- Data Logging
- Data Transfer
- Contextual Help
To know HI5321 HANNA price or buy HI5321 EC / TDS / Salinity and Temperature , please contact Setare Arsh Aria company.
EC Specifications
EC RangeEC ResolutionEC Accuracy (@25ºC/77ºF)EC Calibration
0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC* |
0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm |
±1% of reading (±0.01 μS/cm) |
automatic standard recognition (0.000 μS/cm, 84.00 μS/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration |
TDS Specifications
TDS RangeTDS ResolutionTDS Accuracy (@25ºC/77ºF)
0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor) |
0.001 ppm, 0.01 ppm, 0.1 ppm, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt |
±1% of reading (±0.01 ppm) |
Resistivity Specifications
Resistivity RangeResistivity ResolutionResistivity Accuracy
1.0 to 99.9 Ω•cm; 100 to 999 Ω•cm; 1.00 to 9.99 KΩ•cm; 10.0 to 99.9 KΩ•cm; 100 to 999 KΩ•cm; 1.00 to 9.99 MΩ•cm; 10.0 to 100.0 MΩ•cm |
0.1 Ω•cm; 1 Ω•cm; 0.01 KΩ•cm; 0.1 KΩ•cm; 1 KΩ•cm; 0.01 MΩ•cm; 0.1 MΩ•cm* |
±2% of reading (±1 Ω•cm) |
Salinity Specifications
Salinity RangeSalinity ResolutionSalinity Accuracy (@25ºC/77ºF)Salinity Calibration
practical scale: 0.00 to 42.00 psu; natural seawater scale: 0.00 to 80.00 ppt; percent scale: 0.0 to 400.0% |
0.01 for practical scale/natural seawater scale; 0.1% for percent scale |
±1% of reading |
percent scale—one-point (with HI7037 standard) |
Temperature Specifications
Temperature RangeTemperature ResolutionTemperature AccuracyTemperature CompensationTemperature CoefficientReference Temperature
-20.0 to 120.0 ºC**; -4.0 to 248.0 ºF**; 253.15 to 393.15 K** |
0.1 °C, 0.1 °F, 0.1 K |
±0.2 °C; ±0.4 °F; ±0.2 K (without probe) |
disabled, linear and non-linear (natural water) |
0.00 to 10.00 %/°C |
5.0 to 30.0 °C |
Additional Specifications
Cell ConstantCell TypeElectrode/ProbeProfilesUSP ⟨645⟩ CompliantGLPLoggingDisplayConnectivityEnvironmentPower SupplyDimensionsWeightOrdering InformationNotes
0.0500 to 200.00 |
4 cells |
HI76312 platinum, four-ring EC/TDS probe with and 1 m (3.3’) cable (included) |
up to 10, 5 each channel |
yes |
cell constant, reference temperature, temperature coefficient, calibration points, calibration time stamp, and probe offset for conductivity |
record: 100,000 data point storage, up to 100 lots with max. 50,000 records/lot; interval: settable between 1 second and max log time of 180 minutes; type: automatic, manual, AutoHOLD; additional: 200 records USP |
color graphic LCD with on-screen help, graphing, and custom color configuration |
USB |
0 to 50°C (32 to 122°F; 273 to 323 K), RH max 95% non-condensing |
12 VDC adapter (included) |
160 x 231 x 94 mm (6.3 x 9.1 x 3.7”) |
1.2 kg (2.64 lbs.) |
HI5321 is supplied with HI76312 conductivity probe, HI76404W electrode holder, 12 VDC power adapter and instructions |
*Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation. **Reduced to actual probe limits |