Description
HI83306 Multiparameter Photometer
HI83306 Multiparameter Photometer
HI83306 is Multiparameter Photometer with modern and optical design and suit device for environmental laboratory .
HI83306 works with photometric method.
HI83306 measurs pH via an optional Ph electrode.
HI83306
HI83306
HI83306 measure 16 key water quality parameters with 23 different programmed methods.
HI83306 is useful device to measure Environmental parameters include nitrate, phosphorous and dissolved oxygen.
Key features of HI83300 been listed as below:
- Backlit 128 x 64 Pixel Graphic Liquid Crystal Display
- Built-in Reaction Timer for Photometric Measurements
- Absorbance mode – concentration versus absorbance
- Units of Measurement plus chemical form displayed
- Result conversion at the touch of a button
- Cuvette cover
- pH and temperature measurement with a single probe
- Good Laboratory Practice (GLP) – track calibration information including date, time, buffers used, offset and slope for traceability
- pH CAL Check alerts user to potential problems during the calibration process
- Data Logging – Up to 1000 photometric and pH readings can be stored.
- Logged readings can be quickly and easily transferred to a flash drive or to a PC. Data is exported as a .CSV file for use with spreadsheet programs.
- Battery Status Indicator
- Error Messages
HI83306 characteristics been listed as below:
- High Efficiency LED Light Source
- High Quality Narrow Band Interference Filters
- Reference Detector for a Stable Light Source
- Focusing Lens for Greater Light Yield
HI83306 Multiparameter Photometer Specifications
Input Channels |
|
1 pH electrode input and 5 photometer wavelengths |
pH Electrode |
|
Digital pH electrode (not included) |
Logging type |
|
Log on demand with user name and sample ID optional input |
Logging Memory |
|
1000 readings |
Connectivity |
|
USB-A host for flash drive; micro-USB-B for power and computer connectivity |
GLP |
|
Calibration data for connected pH electrode |
Display |
|
128 x 64 pixel LCD with backlight |
Battery Type/Life |
|
3.7 VDC Li-polymer rechargeable battery/ >500 photometric measurements or 50 hours of continuous pH measurement |
Power supply |
|
5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
Environment |
|
0 to 50.0°C (32 to 122.0°F; 0 to 95% RH, non-condensing |
Dimensions |
|
206 x 177 x 97mm (8.1 x 7.0 x 3.8″) |
Weight |
|
1.0 kg (2.2lbs) |
Photometer/Colorimeter Light Source |
|
5 LEDs with 420nm, 466nm, 525nm, 575nm, and 610nm narrow band interference filters |
Photometre/Colorimeter Light Detector |
|
Silicon Photodetector |
Bandpass filter Bandwidth |
|
8nm |
Bandpass filter wavelength accruacy |
|
+-1nm |
Cuvette Type |
|
Round, 24.6mm |
Number of Methods |
|
128 Max |
Ordering Information |
|
HI-83306 is supplied with sample cuvettes and caps (x4), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual. |
AMMONIA |
|
|
Ammonia Range |
|
Low Range: 0.00 to 3.00 mg/L (as NH3-N) Medium Range: 0.00 to 10.00 mg/L (as NH3-N) High Range: 0.0 to 100.0 mg/L (as NH3-N) |
Ammonia Resolution |
|
0.01 mg/L; 0.1 mg/L |
Ammonia Accuracy |
|
Low Range: ±0.04 mg/L ±4% of reading Medium Range: ±0.05 mg/L ±5% of reading High range: ±0.5 mg/L ±5% of reading at 25 °C |
Ammonia Method |
|
Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method |
CHLORINE |
|
|
Free Chlorine Range |
|
0.00 to 5.00 mg/L (as Cl2) |
Free Chlorine Resolution |
|
0.01 mg/L |
Free Chlorine Accuracy |
|
±0.03 mg/L ±3% of reading |
Total Chlorine Range |
|
0.00 to 5.00 mg/L (as Cl2) |
Total Chlorine Resolution |
|
0.01 mg/L |
Total Chlorine Accuracy |
|
±0.03 mg/L ±3% of reading |
Chlorine Method |
|
Adaptation of the EPA 330.5 DPD method |
CHROMIUM, HEXAVALENT |
|
|
Chromium, Hexavalent Range |
|
Low Range: 0 to 300 μg/L (as Cr6+) High Range: 0 to 1000 μg/L (as Cr6+) |
Chromium, Hexavalent Resolution |
|
1 μg/L |
Chromium, Hexavalent Accuracy |
|
Low Range: ±1 μg/L ±4% of reading High Range: ±5 μg/L ±4% of reading |
Chromium, Hexavalent Method |
|
Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. |
COLOUR, WATER |
|
|
Colour, Water Range |
|
0 to 500 PCU (Platinum Cobalt Units) |
Colour, Water Resolution |
|
1 PCU |
Colour, Water Accuracy |
|
±10 PCU ±5% of reading |
Colour, Water Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method. |
COPPER |
|
|
Copper Range |
|
Low Range: 0.000 to 1.500 mg/L (as Cu2+) High range: 0.00 to 5.00 mg/L (as Cu2+) |
Copper Resolution |
|
0.001 mg/L; 0.01 mg/L |
Copper Accuracy |
|
Low Range: ±0.01 mg/L ±5% of reading High Range ±0.02 mg/L ±4% of reading |
Copper Method |
|
Adaptation of the EPA bicinchoninate method |
CYANURIC ACID |
|
|
Cyanuric Acid Range |
|
0 to 80 mg/L (as CYA) |
Cyanuric Acid Resolution |
|
1 mg/L |
Cyanuric Acid Accuracy |
|
±1 mg/L ±15% of reading |
Cyanuric Acid Method |
|
Adaptation of the turbidimetric method |
MOLYBDENUM |
|
|
Molybdenum Range |
|
0.0 to 40.0 mg/L (as Mo6+) |
Molybdenum Resolution |
|
0.1 mg/L |
Molybdenum Accuracy |
|
±0.3 mg/L ±5% of reading |
Molybdenum Method |
|
Adaptation of the mercaptoacetic acid method |
NICKEL |
|
|
Nickel Range |
|
Low Range: 0.000 to 1.000 mg/L (as Ni) High Range: 0.00 to 7.00 g/L (as Ni) |
Nickel Resolution |
|
0.001 mg/L; 0.01 g/L |
Nickel Accuracy |
|
Low range: ±0.010 mg/L ±7% of reading High Range: ±0.07 g/L ±4% of reading |
Nickel Method |
|
Low Range: Adaptation of the PAN method High Range: Adaptation of the photometric method |
NITRATE |
|
|
Nitrate Range |
|
0.0 to 30.0 mg/L (as NO3– N) |
Nitrate Resolution |
|
0.1 mg/L |
Nitrate Accuracy |
|
±0.5 mg/L ±10% of reading |
Nitrate Method |
|
Adaptation of the cadmium reduction method |
NITRITE |
|
|
Nitrite Range |
|
Freshwater High Range: 0 to 150 mg/L (as NO2-) |
Nitrite Resolution |
|
Freshwater: 1 mg/L |
Nitrite Accuracy |
|
Freshwater High Range: ±4 mg/L ±4% of reading |
Nitrite Method |
|
HR – Adaptation of the Ferrous Sulphate method |
OXYGEN, DISSOLVED |
|
|
Oxygen, Dissolved Range |
|
0.0 to 10.0 mg/L (as O2) |
Oxygen, Dissolved Resolution |
|
0.1 mg/L |
Oxygen, Dissolved Accuracy |
|
±0.4 mg/L ±3% of reading |
Oxygen, Dissolved Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method |
pH |
|
|
pH Range |
|
Photometer: 6.5 to 8.5 pH pH electrode: -2.00 to 16.00 pH |
pH Resolution |
|
Photometer: 0.1 pH pH electrode: 0.1 pH |
pH Accuracy |
|
Photometer: ±0.1 pH pH electrode: ±0.01 pH |
pH Calibration |
|
Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
pH Temperature Compensation |
|
Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used |
pH CAL Check (electrode diagnostics) |
|
clean electrode and check buffer/check probe displayed during calibration |
pH Method |
|
Photometer: phenol red |
pH-mV Range |
|
±1000 mV |
pH-mV Resolution |
|
0.1 mV |
pH-mV Accuracy |
|
±0.2 mV |
PHOSPHATE |
|
|
Phosphate Range |
|
Freshwater Low Range: 0.00 to 2.50 mg/L (as PO43-) High range: 0.0 to 30.0 mg/L (as PO43-) |
Phosphate Resolution |
|
Freshwater: 0.01 mg/L; 0.1 mg/L |
Phosphate Accuracy |
|
Freshwater Low Range: ±0.04 mg/L ±4% of reading High Range: ±1 mg/L ±4% of reading |
Phosphate Method |
|
Freshwater Low Range: Adaptation of the Ascorbic Acid method Freshwater High Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method |
SILICA |
|
|
Silica Range |
|
Low Range: 0.00 to 2.00 mg/L (as SiO2) |
Silica Resolution |
|
0.01 mg/L |
Silica Accuracy |
|
Low Range: ±0.03 mg/L ±3% of reading |
Silica Method |
|
Low Range: Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue method |
SILVER |
|
|
Silver Range |
|
0.000 to 1.000 mg/L (as Ag) |
Silver Resolution |
|
0.001 mg/L |
Silver Accuracy |
|
±0.020 mg/L ±5% of reading |
Silver Method |
|
Adaptation of the PAN method |
ZINC |
|
|
Zinc Range |
|
0.00 to 3.00 mg/L (as Zn) |
Zinc Resolution |
|
0.01 mg/L |
Zinc Accuracy |
|
±0.03 mg/L ±3% of reading |
Zinc Method |
|
Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method |
REAGENTS |
|
|
Ammonia HR, MR & LR |
|
HR – HI-93733-01 (100 tests); HI-93733-03 (300 tests); MR – HI-93715-01 (100 tests); HI-93715-03 (300 tests); LR – HI-93700-01 (100 tests); HI-93700-03 (300 tests) |
Chlorine – Free |
|
Free Chlorine Reagents: HI-93701-01 (100 tests); HI-93701-03 (300 tests); HI-93701-F (liquid) (300 tests) |
Chlorine Total |
|
HI-93711-01 (100 tests); HI-93711-03 (300 tests); HI-93711-T (300 tests liquid)
Chlorine, Total ULR: HI-95761-01 (100 tests); HI-95761-03 (300 tests) |
Chromium VI – HR & LR |
|
HR – HI-93723-01 (100 tests); HI-93723-03 (300 tests)
LR – HI-93749-01 (100 tests); HI-93749-03 (300 tests) |
Copper – HR & LR |
|
HR – HI-93702-01 (100 tests); HI-93702-03 (300 tests)
LR – HI-95747-01 (100 tests); HI-95747-03 (300 tests) |
Cyanuric Acid |
|
HI-93722-01 (100 tests); HI-93722-03 (300 tests) |
Molybdenum |
|
HI-93730-01 (100 tests); HI-93730-03 (300 tests) |
Nickel – HR & LR |
|
HR – HI-93726-01 (100 tests); HI-93726-03 (300 tests)
LR – HI-93740-01 (50 tests); HI-93740-03 (150 tests) |
Nitrate |
|
HI-93728-01 (100 tests); HI-93728-03 (300 tests) |
Nitrite – HR & LR |
|
HR – HI-93708-01 (100 tests); HI-93708-03 (300 tests)
LR – HI-93707-01 (100 tests); HI-93707-03 (300 tests) |
Oxygen, Dissolved (DO) |
|
HI-93732-01 (100 tests); HI-93732-03 (300 tests) |
pH |
|
HI-93710-01 (100 tests); HI-93710-03 (300 tests) |
Phosphate – HR & LR |
|
HR – HI-93717-01 (100 tests); HI-93717-03 (300 tests)
LR – HI-93713-01 (100 tests); HI-93713-03 (300 tests) |
Silica – LR |
|
LR – HI-93705-01 (100 tests); HI-93705-03 (300 tests) |
Silver |
|
HI-93737-01 (50 tests); HI-93737-03 (150 tests) |
Zinc |
|
HI-93731-01 (100 tests); HI-93731-03 (300 tests) |
To know HI83306 HANNA price or buy HI83306 Multiparameter Photometer ,please contact Setare Arsh Aria company.
Reviews