HI98703 Turbidity Meter
The HI98703 is a portable Turbidity meter with high accuracy at low range (below 0.05 NTU).
HI98703 has Two, three or four point calibration and GLP features .
HI98703 can log up to 200 records and is very user friendly .
- The HI 98703 meets and exceeds the requirements of the USEPA Method 180.1 for wastewater and Standard Method 2130 B for drinking water.
delivery scope of HI98703 : HI 98703-02 (230V) is supplied with sample cuvettes and caps (5), HI 98703-11 calibration cuvettes, HI 93703-58 silicone oil, cuvette cleaning cloth, batteries, AC adapter, instruction manual and rugged carrying case.
To know HI98703 HANNA price or buy HI98703 Turbidity Meter ,please contact Setare Arsh Aria company.
HI98703 Turbidity Meter Specification :
Range | 0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU | |
Range Selection | automatic | |
Resolution | 0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU from 10.0 to 99.9 NTU; 1 NTU from 100 to 1000 NTU | |
Accuracy @25°C | ±2% of reading plus 0.02 NTU | |
Repeatability | ±1% of reading or 0.02 NTU, whichever is greater | |
Stray Light | < 0.02 NTU | |
Light Detector | silicon photocell | |
Light Source | tungsten filament lamp | |
Lamp Life | greater than 100,000 readings | |
Method | Ratio Nephelometric Method (90°), ratio of scattered and transmitted light; Adaptation of the USEPA Method 180.1 and Standard Method 2130 B | |
Measuring Mode | normal, average, continuous | |
Turbidity Standards | <0.1, 15, 100 and 750 NTU | |
Calibration | two, three or four-point calibration | |
Log Memory | 200 records | |
PC Connectivity | USB or RS232 | |
Environment | up to 50°C (122°F); RH max 95% non-condensing | |
Power Supply | 1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use | |
Dimensions / Weight | 224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) / 512 g (18 oz.) |